Hepo Adult Portable Bed Rails Recall — HIGH Severity
Mobility Source Medical Technology Co. · Recalled April 30, 2026 · Furniture
Data source: U.S. Consumer Product Safety Commission (CPSC) Last verified: May 1, 2026
Product Description
This recall involves Hepo-branded adult portable bed rails, model 110039. The recalled bed rails have white tubing with black foam rubber handle grips, a black fabric pouch, and measures about 16 inches wide by 25 inches high. The model number is located on the original product packaging.
Hazard
The recalled bed rails violate the mandatory standard for adult portable bed rails because users can become entrapped within the bed rail or between the bed rail and the side of the mattress, posing a serious entrapment hazard and risk of death by asphyxiation. The bed rails also do not meet structural stability or retention strap requirements, posing a fall hazard. The bed rails' push pins and push pin holes are incorrectly sized, posing a laceration hazard. In addition, the bed rails do not bear the required hazard warning labels.
Reported Injuries
No injuries have been reported as of the recall date.
What You Should Do
Consumers should stop using the recalled adult portable bed rails immediately and contact Hepo Care Medical Equipment Online for a full refund. Consumers will be asked to destroy the bed rails by dismantling the bed rails, cutting off the storage bag and writing "RECALLED" on the upper and lower rails with permanent marker, take a photo of the destroyed rails and upload the photo to www.hepolife.com/bed-rail-safety-recall or email the photo to [email protected].
Manufacturer Contact
Hepo Care Medical Equipment Online by email at [email protected] or online at https://www.hepolife.com/bed-rail-safety-recall or https://www.hepolife.com and click "Recall" at the top of the page for more information.
Recall Details
Units sold: About 2,200
Sold at: Online at Walmart.com from November 2025 through January 2026 for about $100 Amazon.com from June 2025 through January 2026 for between $35 and $40.
Manufactured by: Mobility Source Medical Technology Co., Ltd. of China; Lingqingxiangjiankangkejigufenyouxiangongsi (Lingqingxiang Health Technology Co., Ltd.) dba Hepo Care Medical Equipment Online Ltd., of China
Country of origin: China
Source